8.8 Hold outputs
(PATH: Menu/Configure/Hold Outputs)
This Menu allows for the configuration of the Hold status. The HOLD function is used if a Easy
Clean system is controlled from the M300 or if a control of the HOLD functionality is required
from an overlying system.
Initiating a hold condition will maintain the analog signal output and relay status at the value/
state at the time the Hold is initiated, for as long as the Hold state is maintained. In addition,
the USB output will be held if the USB Hold option is set to "Last Values". The USB Hold feature
is set to "Off" by default. Refer to section 9.2 for more information on the USB settings.
Analog outputs, Relays and USB will not be held if No is selected. If Yes is selected, Outputs will
be held depending of the status of Digital Input # 1. The choices are 'high' or 'low'. All outputs
will be held if the digital input is in the selected state. If 'Off' is selected as DI (Digital Input)
status, the Digital Input is inactive and the Hold status will not be triggered through an external
signal, although the outputs will be held during configuration or calibration procedures as long
as the hold Outputs option is 'Yes.