M300 32
© 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter O
2
M300
Printed in Switzerland 52 121 305
8.3.3 O
2
Press the [ENTER] key to select this Menu, which allows for the setting of additional
measurement parameters for the O
2
measurement.
Atmospheric pressure and Process Pressure can be set. The default value for AtmPres is 760.0
and the default unit is mmHg. The units ProcPres and AtmPres do not have to be the same.
The salinity of the measured solution and the relative humidity of the calibration gas can also
be entered. The allowed values for Relative Humidity are in the range 0.00 to 1.00. Pressing
the [ENTER] key again will bring up the Save Changes dialog.
Selecting "No" will discard the entered values, selecting „Yes" will make the entered values the
current ones.
8.3.4 Set averaging
Press the [ENTER] key to select this Menu. The averaging method (noise filter) for each channel
can now be selected. The options are Special (Default), None, Low, Medium and High:
None = no averaging or filtering
Low = equivalent to a 3 point moving average
Medium = equivalent to a 5 point moving average
High = equivalent to a 7 point moving average
Special = averaging depending on signal change (normally High averaging but Low averaging
for large changes in input signal)
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting „No" will
discard the entered values, selecting „Yes" will make the entered values the current ones.