8.6.2 Clean
Configure the Relay to be used for the cleaning cycle.
The Cleaning Interval can be set from 0.000 to 999.9 hours. Setting it to 0 turns the clean
cycle is off. The cleaning time can be 0 to 9999 seconds and must be smaller than the
Cleaning Interval.
Select the desired Relay state: Normal or Inverted.
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting „No" will
discard the entered values, selecting „Yes" will make the entered values the current ones.
8.7 Display
(PATH: Menu/Configure/Display)
This Menu allows for the configuration of the values to be displayed and also the configuration
of the Display itself.
8.7.1 Measurement
The Display has in 4 Lines. Line 1 on top and Line 4 on the bottom.
Select the values (channels) to be displayed on each line of the display.
If this is set to On then when an alarm occurs the message "Failure – Press [ENTER] will be
on Line 4 in the normal measurement mode.
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting „No" will
discard the entered values, selecting „Yes" will make the entered values the current ones.