Transmitter O
2
M300 24
© 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter O
2
M300
Printed in Switzerland 52 121 305
6 Quick setup
(PATH: Menu/Quick Setup)
While in Measurement mode press the [MENU] key to bring up the Menu selection. Select
"Quick Setup" and press the [ENTER] key. The display prompts you to enter the Quick Setup
security code. Enter the Quick Setup security code to "xxxxx", then press the [ENTER] key to
confirm the Quick Setup security code.
Convention:
1
st
line on display => a
2
nd
line on display => b
3
rd
line on display => c
4
th
line on display => d
Only lines a and b can be configured in Quick setup. Go to the Configuration Menu to configure
lines c and d.
6.1 Channel selection
Select a or c to configure the values to be displayed. Please use the Configuration menu to freely
configure the channels.
Example:
By selecting a and O
2
as unit, the O
2
value will be displayed on the 1
st
line. By selecting b and
O
2
as unit, the O
2
value will be displayed on the 3
rd
line of the display.
By selecting a and °C as unit, the temperature will be displayed on the 1
st
line. By selecting b
and °C as unit, the temperature will be displayed on the 3
rd
line.
6.2 Analog outputs
On the screen above, by selecting „Yes" the linear 4-20 mA analog output Aout1 will be setup
when [ENTER] is pressed. Selecting „No" means that no analog output is setup.
Aout1 min, Aout1 max are the minimum and maximum measurement values for the 4 and
20mA values respectively. To configure output 2, use the Configuration Menu. Press [ENTER}.