7.4 Temperature calibration
Enter Calibration mode as described in section 7.2 and select "Temperature Calibration".
7.4.1 One – point temperature calibration
See 7.4 on how to enter the Temperature Calibration menu. Slope or Offset can be selected with
the 1 point calibration. Select "Slope" to recalculate the Slope factor M and Offset to recalculate
the offset calibration factor A.
Enter Point 1
The newly calculated value – either M or A – is displayed. Select „Yes" to save the new
calibration values and the successful Calibration is confirmed on the display.
7.4.2 Two – point temperature calibration
See 7.4 on how to enter the Temperature Calibration menu. Select 2 point as calibration Type.
Enter the value for Point 1 and press [ENTER].
Enter the value for Point 2 and press [ENTER].