Transmitter Multiparameter M400 71
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400
Printed in Switzerland 52 121 378
13 Maintenance
13.1 Front Panel Cleaning
Clean the front panel with a damp soft cloth (water only, no solvents). Gently wipe the surface
and dry with a soft cloth.
14 Troubleshooting
If the equipment is used in a manner not specified by Mettler-Toledo Thornton, Inc.,
the protection provided by the equipment may be impaired.
Review the table below for possible causes of common problems:
Problem Possible Cause
Display is blank.
No power to M400.
Blown fuse.
LCD display contrast set incorrectly.
Hardware failure.
Incorrect measurement readings.
Sensor improperly installed.
Incorrect units multiplier entered.
Temperature compensation incorrectly set or
disabled.
Sensor or transmitter needs calibration.
Sensor or patch cord defective or exceeds
recommended maximum length.
Hardware failure.
Measurement readings not stable.
Sensors or cables installed too close to equipment
that generates high level of electrical noise.
Recommended cable length exceeded.
Averaging set too low.
Sensor or patch cord defective.
Displayed is flashing. Setpoint is in alarm condition (setpoint exceeded).
Cannot change menu settings. User locked out for security reasons.