9.4 Set / Clear Lockout
(PATH: Menu / System / Set / Clear Lockout)
This menu enables / disables the Lockout functionality of the transmitter. The user will be asked
for a password before being allowed into any menus if the Lockout functionality is enabled.
The Lockout – Menu is protected: Enter the Administrator or Operator Password and select YES to
enable or NO to disable the Lockout functionality. Pressing the [ENTER] key after the selection
will bring up the Save Changes dialog. Selecting No will discard the entered value, selecting Yes
will make the entered value the current one.
9.5 Reset
(PATH: Menu / System / Reset)
This Menu allows access to the following options:
Reset System, Reset Meter Cal, Reset Analog Cal.
9.5.1 Reset System
This Menu allows the reset of the meter to the factory default settings (Setpoints off, analog
outputs off, etc.). The meter calibration and the analog output calibration are not affected.
Pressing the [ENTER] key after the selection will bring up a confirmation screen. Selecting No
will return the user to the Measurement mode with no changes. Selecting Yes will reset the meter.
9.5.2 Reset Meter Calibration
This Menu allows the reset of the meter’s calibration factors to the last factory calibration values.
Pressing the [ENTER] key after the selection will bring up a confirmation screen. Selecting No
will return the user to the Measurement mode with no changes. Selecting Yes will reset the meter
calibration factors.