Transmitter Multiparameter M400 23
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400
Printed in Switzerland 52 121 378
4.4.5 Typical Wiring (Using TB3 / TB4)
4.4.5.1 Example 1
pH measurement without Solution Ground
Cable
Combination
pH electrode
Temperature
probe
yellowgreen
green
red
transparent
white
Jumper
1
2
3
4
5
6
7
8
9
C
NOTE: Jumper Terminals 3 and 4.
Wire colors only valid for connection with VP cable; blue and grey not connected.
1 – Glass
2 – Not used
3 – Reference
4 – Shield / GND
5 – Not used
6 – Solution GND / RTD ret
7 – Not used
8 – RTD
9 – Not used