Transmitter Multiparameter M400 66
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400
Printed in Switzerland 52 121 378
11.2.1.4 Rg Diagnostic
Rg Diagnostic is performed as a two point calibration. Navigate to the Calibrate Meter screen
and select Channel A and Rg Diagnostic.
Enter the value for Point 1 of the calibration according to the resistor connected across the
pH glass electrode measuring input. Press [ENTER] to begin the calibration process.
Enter the value for Point 2 of the calibration according to the resistor connected across the
pH glass electrode measuring input.
Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Select Yes
to save the calibration values and the successful Calibration is confirmed on the display. The
transmitter will return to the Measurement mode in approximately 5 seconds.
11.2.1.5 Rr Diagnostics
Rr Diagnostic is performed as a two point calibration. Navigate to the Calibrate Meter screen
and select Channel A and Rr Diagnostic.
Enter the value for Point 1 of the calibration according to the resistor connected across the
pH reference measuring input. Press [ENTER] to begin the calibration process.
Enter the value for Point 2 of the calibration according to the resistor connected across the
pH reference measuring input.
Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Select Yes
to save the calibration values and the successful Calibration is confirmed on the display. The
transmitter will return to the Measurement mode in approximately 5 seconds.