Transmitter Multiparameter M400 58
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400
Printed in Switzerland 52 121 378
10.3 Mode
(PATH: MENU / PID Setup / Mode)
This menu contains the selection of control modes using relays or analog outputs. Press
[ENTER].
10.3.1 PID Mode
This menu assigns a relay or analog output for PID control action as well as details of their
operation. Based on the control device being used, select one of the following three paragraphs
for use with solenoid valve, pulse input metering pump or analog control.
Pulse Length – If using a solenoid valve, select ”Relays” and ”PL”, Pulse Length.
Choose the first relay position as #3 (recommended) and / or the second relay position
as #4 (recommended) as well as the Pulse Length (PL) according to the table below. A longer
pulse length will reduce wear on the solenoid valve. The % ”on” time in the cycle is proportional
to the control output.
C
NOTE: All relays from #1 to #6 can be used for the controlling function.
1
st
Relay Position
(#3)
2
nd
Relay Position
(#4)
Pulse Length (PL)
Conductivity
Controlling
concentrating reagent
feed
Controlling dilution water
Short (PL) provides
more uniform feed.
Suggested start point
= 30 sec
pH / ORP
Feeding base Feeding acid
Reagent addition cycle:
short PL provides more
uniform addition of
reagent. Suggested start
point = 10 sec
Dissolved
Oxygen
Reverse control action
Direct acting control
action
Feed cycle time: short
PL provides more
uniform feed. Suggested
start point = 30 sec