Transmitter M400/2(X)H, M400G/2XH 82
© 03 / 2014 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M400/2(X)H, M400G/2XH
Printed in Switzerland 30 031 683
12 Info
(PATH: Info)
Info
Calibration
Data
Messages
Model/Software
Revision
ISM Sensor
Info*
ISM
Diagnostics*
* Only available in combination
with ISM sensors
Pressing the . key will display the Info menu with the options Messages, Calibration Data and
Model/Software Revision.
12.1 Messages
(PATH: Info/Messages)
The most recent message is displayed. The up and down arrow keys allow scrolling through the
last four messages that have occurred.
Clear Messages clears all the messages. Messages are added to the message list when the
condition that generates the message first occurs. If all messages are cleared and a message
condition still exists and started before the clear then it will not appear in the list. For this mes-
sage to re-occur in the list the condition must go away and then reappear.
Press [ENTER] to exit from this display.
12.2 Calibration data
(PATH: Info/Calibration Data)
Selecting Calibration Data displays the calibration constants for each sensor.
P = calibration constants for the primary measurement
S = calibration constants for the secondary measurement
Press . for ORP calibration data of ISM pH sensors.
Press [ENTER] to exit from this display.