8.2.2 Temperature source (only for analog sensors)
(PATH: Menu/Configure/Measurement/Temperature Source)
Enter Measurement as described in chapter 8.2 “Measurement”. Select Temperature Source by
using the or . key and press [ENTER].
The following options can be chosen:
Auto: The transmitter automatically recognizes the temperature source.
Use NTC22K: Input will be taken from the sensor attached.
Use Pt1000: Temperature input will be taken from the sensor attached
Use Pt1 00: Input will be taken from the sensor attached.
Fixed = 25 °C: Allows a specific temperature value to be entered. It must be chosen when
customer use pH sensor without temperature source.
h
NOTE: If temperature source is set to Fixed, the temperature applied during one-point and/or two-
point calibration of pH electrodes can be adjusted within the corresponding calibration proce-
dure. After the calibration the fixed temperature defined in this configuration menu is valid again.
Pressing the [ENTER] key will bring up the Save Changes dialog.
Selecting No will discard the entered values and return to the measurement display screen, se-
lecting Yes will save changes made.
8.2.3 Parameter related settings
(PATH: Menu/Configure/Measurement/pH)
Additional measurement and calibration parameters can be set for each parameter; conductivity,
pH, and O
2
.
h
NOTE: Use pH menu for settings of pH/pNa sensors.
Enter Configuration Mode as described in section 8.1 “Enter Configuration mode” and select the
menu Measurement (see section 8.2 “Configuration/Measurement”).
Depending on the connected sensor, the menu pH, O2, can be selected by using the A or .
key. Press [ENTER]
For more details, please see the following explanations depending on the selected parameter.