Transmitter M400/2(X)H, M400G/2XH 76
© 03 / 2014 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M400/2(X)H, M400G/2XH
Printed in Switzerland 30 031 683
11.1.3 Display
All pixels of the display will be lit for 15 seconds to allow troubleshooting of the display. After
15seconds the transmitter will return to the normal measuring mode or press [ENTER] to exit
sooner.
11.1.4 Keypad
For keypad diagnostics, the display will indicate which key is pressed. Pressing [ENTER] will re-
turn the transmitter to the normal measuring mode.
11.1.5 Memory
If Memory is selected then the transmitter will perform a RAM and ROM memory test. Test pat-
terns will be written to and read from all RAM memory locations. The ROM checksum will be re-
calculated and compared to the value stored in the ROM.
11.1.6 Set OC
The Set OC diagnostic menu allows to open or close each OC manually. To access OC and 6,
press [ENTER].
0 = open the OC
1 = close the OC
Press [ENTER] to return to Measurement mode.