Mettler Toledo M400/2(X)H Especificaciones Pagina 80

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 111
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 79
Transmitter M400/2(X)H, M400G/2XH 80
© 03 / 2014 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M400/2(X)H, M400G/2XH
Printed in Switzerland 30 031 683
Enter the value for Point 1 of the calibration according to the resistor connected across the pH
glass electrode measuring input. Press [ENTER] to begin the calibration process.
Enter the value for Point 2 of the calibration according to the resistor connected across the pH
glass electrode measuring input.
Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Select Yes to
save the calibration values and the successful calibration is confirmed on the display. The trans-
mitter will return to the measurement mode in approximately 5 seconds.
11.2.1.5 Rr diagnostic
Rr diagnostic is performed as a two point calibration. Navigate to the Calibrate Meter screen and
select Channel A and Rr Diagnostic.
Enter the value for Point 1 of the calibration according to the resistor connected across the pH
reference measuring input. Press [ENTER] to begin the calibration process.
Enter the value for Point 2 of the calibration according to the resistor connected across the pH
reference measuring input.
Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Select Yes to
save the calibration values and the successful calibration is confirmed on the display. The trans-
mitter will return to the measurement mode in approximately 5 seconds.
11.2.1.6 Calibrate analog output signals
Select the Analog Output you wish to calibrate. Each analog output can be calibrated at 4 and
20 mA.
Vista de pagina 79
1 2 ... 75 76 77 78 79 80 81 82 83 84 85 ... 110 111

Comentarios a estos manuales

Sin comentarios