Transmitter M300 FLOW 56
© 09 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M300 FLOW
Printed in Switzerland 52 121 319
11.1.2 Digital Input
The digital Input menu shows the state of the digital input. Press [ENTER] to exit from this
display.
11.1.3 Display
All pixels of the display will be lit for 15 seconds to allow troubleshooting of the display. After
15 seconds the transmitter will return to the normal Measuring mode or press [ENTER] to exit
sooner.
11.1.4 Keypad
For the keypad diagnostics the display will indicate which key is pressed. Pressing [ENTER]
will return the transmitter to the normal Measuring mode.
11.1.5 Memory
If Memory is selected then the transmitter will perform a RAM and ROM memory test. Test
patterns will be written to and read from all RAM memory locations. The ROM checksum will be
recalculated and compared to the value stored in the ROM.
Comentarios a estos manuales