8.2.1 Set Averaging
Press [ENTER] to select this Menu. The averaging method (noise filter) for each measurement
can now be selected. The options are Special (Default), None, Low, Medium and High.
None = no averaging or filtering
Low = equivalent to a 3 point moving average
Medium = equivalent to a 5 point moving average
High = equivalent to a 7 point moving average
Special = averaging depending on signal change (ideal for large changes in input signal).
Pressing the [ENTER] key will scroll through the remaining measurements.
Press[ENTER] to bring up the Save Changes dialog. Selecting No will discard the entered values
and return to the measurement display screen, selecting Yes will save changes made.
8.2.2 Set Pipe ID
Press [ENTER] key to select this Menu. This menu is used for flow sensors where the inside
diameter of the pipe where the flow sensor is installed, is needed to calculate an accurate flow
velocity measurement.
Enter the inside diameter of the pipe. The value can be entered in (inches) or (cm).
Press [ENTER] to configure channels C and D of four channel transmitters.
Press [ENTER]to bring up the Save Changes dialog.