Transmitter M300 FLOW 25
© 09 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M300 FLOW
Printed in Switzerland 52 121 319
The following wiring information is used when connecting Spirax Sarco/Emco flow (formerly
Fluidyne) insertion type flow sensors. Thornton models 33358 thru 33375.
THORNTON 33358-33375
Spirax Sarco/Emco flow
(formerly Fluidyne)
Insertion Vortex Sensor
Type: high
M300 transmitter
(–)
GND
(+)
1K
0.33uF
TB 3
1
2 GND
3
4
5 +5V
6 GND
7 A J
8 A IN
9 +5V
Extension cable not provided. Use 2-conductor twisted pair with shield, 20 AWG (Belden 9320
or equivalent), 2000 ft (610 m) maximum length.
4.7.2 Wiring for ”LOW” type flow sensors
The following wiring information is used when connecting (GF Signet 515) type flow sensors.
Thornton models 33189, 33193, 33195, 33196, and 33229.
THORNTON 33189, 33193,
33195, 33196, 33229
Signet 515 Sensor
Type: low
M300 transmitter
black
shield
red
TB 3
1
2 GND
3
4
5 +5V
6 GND
7 A J
8 A IN
9 +5V
Extension cable not provided. Use 2-conductor twisted pair with shield, 22 AWG (Belden 8451