Transmitter M300 FLOW 24
© 09 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M300 FLOW
Printed in Switzerland 52 121 319
The following wiring information is used when connecting Sanitary Turbine type flow sensors.
Thornton models 33336 thru 33377 (Hoffer) and 33441 thru 33450 (Sponsler).
THORNTON
33336-333377 (Hoffer)
Hoffer Turbine Sensors
Type: high
M300 transmitter
black
shield
red
white
0.33uF
TB 3
1
2 GND
3
4
5 +5V
6 GND
7 A J
8 A IN
9 +5V
20 ft (6.1 m) extension cable provided with sensor. Use 3-conductor with shield,
20 AWG (Belden 9364 or equivalent) to extend length to 3000 ft (915 m) maximum.
THORNTON
33441-33450 (Sponsler)
Sponsler Turbine Sensor
Type: high
M300 transmitter
black
shield
red
orange
0.33uF
TB 3
1
2 GND
3
4
5 +5V
6 GND
7 A J
8 A IN
9 +5V
20 ft (6.1 m) extension cable provided with sensor. Use 3-conductor with shield,
20 AWG (Belden 9364 or equivalent) to extend length to 3000 ft (915 m) maximum.