Transmitter M300 60
© 12 / 11 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M300
Printed in Switzerland 52 121 389
In addition the relative humidity of the calibration gas can also be entered. The allowed values
for relative humidity are in the range 0% to 100%.
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will dis-
card the entered values and return to the measurement display screen, selecting Yes will save
changes made.
8.2.5 SetAveraging
(PATH: Menu/Configure/Measurement/Set Averaging)
Press the [ENTER] key to select this Menu. The averaging method (noise filter) for each mea-
surement line can now be selected. The options are Special (Default), None, Low, Medium and
High:
None = no averaging or filtering
Low = equivalent to a 3 point moving average
Medium = equivalent to a 6 point moving average
High = equivalent to a 10 point moving average
Special = averaging depending on signal change (normally High averaging but Low averaging
for large changes in input signal)
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard
the entered values and return to the measurement display screen, selecting Yes will save chang-
es made.
8.3 AnalogOutputs
(PATH: Menu/Configure/Analog Outputs)
Enter configuration mode as described in Section 8.1 ”Enter Configuration Mode” and navigate
to the menu “Analog Outputs” by using the m or . key.
Press the [ENTER] key to select this Menu, which lets you configure the 2 (4 for dual channel
version) Analog Outputs.
Once analog outputs have been selected, use the
b and c buttons to navigate between config-
urable parameters. Once a parameter is selected, its setting can be selected per the following ta-
ble:
When an Alarm Value is selected, the analog output will go to this value if any alarm condition
occurs.
Parameter Selectable Values
Aout: 1, 2, 3* or 4* (default is 1)
Measurement: a, b, c, d or blank (none) (default is blank)
Alarm Value: 3.6 mA, 22.0 mA or Off (default is off)
* Only on dual channel version.
The Aout type can be Normal, Bi-Linear, Auto-Range or Logarithmic. The range can be 4–20mA
or 0–20mA. Normal provides linear scaling between the minimum and maximum scaling l