Transmitter M300 39
© 12 / 11 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M300
Printed in Switzerland 52 121 389
7 SensorCalibration
(PATH: Cal)
The calibration key c allows the user one-touch access to Sensor calibration and verification
features. Thornton models also allow access to Meter and Analog Output calibration (see chap-
ter 11.3.1 and 11.3.2). All other models also allow access to Analog Output and Meter calibra-
tion if the access has been previously unlocked (see chapter 11.3.3 ”Calibrate Unlock”).
h
NOTE: During Calibration, a flashing ”H” in the upper left corner of the display indicates a
calibration is in process with a Hold condition active. (The hold output function need to be
activated.) See also chapter 3.3 “Display”
7.1 EnterCalibrationMode
While in Measurement mode press the c key. If the display prompts you to enter the calibration
security code, press the m or . key to set the calibration security mode, then press the [EN-
TER] key to confirm the calibration security code.
For dual-channel devices: Using the the m or . key on the ”Channel A” field let you change the
channel to be calibrated. Then use the c key to move to the calibration field.
Press the m or . key to select the type of calibration desired. The choices for each sensor type
are:
Conductivity = Conductivity, Resistivity, Temperature*, Edit*, Verify
Oxygen** = Oxygen, Temperature*, Edit*, Verify
Ozone** = Ozone, Temperature*, Edit*, Verify
pH = pH, mV, Temperature*, Edit pH*, Edit mV, Verify, ORP***
Press [ENTER].
* not on ISM version
** for Thornton transmitters (p/n 58 001 316 and 58 001 306), a jumper is required between
terminal 1, 3 and 4 on TB3 and/or TB4.
*** only with ISM sensors
After every successful calibration, the following options are available:
Calibrate: Calibration values will be overtaken und used for the measurement. Additionally,
the data will be stored in the sensor*.
Abort: Calibration values will be discarded.
* only available with ISM sensors