
Transmitter M300 45
© 12 / 11 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M300
Printed in Switzerland 52 121 389
7.4.1 One-PointSensorCalibration
Enter Ozone Calibration mode as described in section 7.1 “Enter Calibration Mode” and select
Ozone.
Ozone sensor calibration is always either a one point Comparison (Slope) or a Zero (Offset) cal-
ibration. A one point slope calibration is always obtained from a comparison instrument
or colorimetric test kit and a one point offset calibration is done in air or in Ozone-free water.
Select 1 point followed by either Slope or ZeroPt as the calibration type. Press [ENTER].
Enter the value for Point 1 including a decimal point and units. The value in the second text line
is the value being measured by the transmitter and sensor in the units selected by the user.
Press [ENTER] when this value is stable to perform the calibration.
After the calibration the slope calibration factor S and the offset calibration factor Z are displayed.
Select Yes to save the calibration values and the successful Calibration is confirmed on the dis-
play.
You will get the message ”Re-install sensor” and ”Press Enter” on the display. After pressing
”Enter” the M300 returns to the measuring mode.
7.5 pHCalibration
For pH sensors, the M300 transmitter features one-point, two-point (Auto or Manual mode)
or process calibration with 9 preset buffer sets or manual buffer entry. Buffer values refer to
25 °C. To calibrate the instrument with automatic buffer recognition, you need a standard pH
buffer solution that matches one of these values (see section 8.2.4.2 ”pH Parameters” for con-
figuring modes and selecting buffer sets). Please select the correct buffer table before using au-
tomatic calibration (see chapter 20 “Buffer tables”).
h
NOTE: Only buffer table 20.2.1 Mettler-pH/pNa is available for dual membrane pH electrodes
(pH/pNa)
Enter pH Calibration mode as described in section 7.1 “Enter Calibration Mode”.
7.5.1 OnePointCalibration
Select 1 point Calibration. With pH sensors a one point calibration is always performed as a off-
set calibraton.
Depending on the parameterized Drift control (see chapter 8.2.4.2 “pH Parameters”) one of the
two following modes is active.
Comentarios a estos manuales