Mettler Toledo M300 ISM Especificaciones Pagina 32

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 122
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 31
Transmitter M300 32
© 12 / 11 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M300
Printed in Switzerland 52 121 389
4.4.4.1 Example4
ORP measurement with pH Solution ground electrode (e.g. InPro 3250 SG, InPro 4800 SG).
1
2
3
5
6
7
8
9
yellowgreen
green
red
transparent
white
Combination
pH electrode
with RTD
and SG
Temperature
probe
blue
Cable
Not connected
4
Jumper
h
NOTE: Jumper Terminal 3 and 4.
h
NOTE:Pt100adapterrequired(enclosed)forsensorswithPt100temperatureprobe.
Forwiringdetailsseepage24.
1: Platinum 6: RTD ret
2: Not used 7: Not used
3: Reference 8: RTD
4: Shield/GND 9: Not used
5: Not used