Transmitter M400 60
© 02 / 2010 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M400
Printed in Switzerland 52 121 378
Select Process Calibration. With ISFET sensors a process calibration is always performed as a
offset calibration.
The following calibration steps are the same as described for pH sensors (see chapter 7.6.3
“Process calibration”).
7.7.4 mV calibration
Enter mV calibration mode as described in section 7.1 “Enter Calibration Mode”.
The following calibration steps are the same as described for pH sensors (see chapter 7.6.4
“mV calibration”).
7.8 Dissolved carbon dioxide calibration
For dissolved carbon dioxide (CO
2
) sensors, the M400 transmitter features one-point,
two-point or process calibration. For the one-point or two-point calibration the solution with
pH = 7.00 and/or pH = 9.21 of the Mettler – 9 standard buffer has to be used (see also chapter
8.2.3.85 “Dissolved carbon dioxide parameters”).
7.8.1 One point calibration
Enter CO
2
calibration mode as described in section 7.1 “Enter Calibration Mode”.
Select 1 point Calibration. With CO2 sensors a one point calibration is always performed as a
offset calibration.
Depending on the parameterized Drift control (see chapter 8.2.3.5 “Dissolved carbon dioxide
parameters”) one of the two following modes is active.
7.8.1.1 Auto Mode
Place the electrode in the buffer solution and press the [ENTER] key to start the calibration.
The display shows the buffer the transmitter has recognized (Point 1) and the measured value.