Mettler Toledo M400 Especificaciones Pagina 33

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 131
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 32
Transmitter M400 33
© 02 / 2010 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M400
Printed in Switzerland 52 121 378
4.5.3.4 Example 4
ORP measurement with pH Solution ground electrode (e.g. InPro 3250 SG, InPro 4800 SG).
1
2
3
5
6
7
8
9
yellowgreen
green
red
transparent
white
Combination
pH electrode
with RTD
and SG
Temperature
probe
blue
Cable
Not connected
4
Jumper
h
NOTE: Jumper terminal 3 and 4
1: Platinum
2: Not used
3: Reference
4: Shield / GND
5: Not used
6: RTD ret
7: Not used
8: RTD
9: Not used