11.2.1.5 Rr diagnostics
Rr diagnostic is performed as a two point calibration. Navigate to the Calibrate Meter screen
and select Channel A and Rr Diagnostic.
Enter the value for Point 1 of the calibration according to the resistor connected across the
pH reference measuring input. Press [ENTER] to begin the calibration process.
Enter the value for Point 2 of the calibration according to the resistor connected across the
pH reference measuring input.
Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Select Yes
to save the calibration values and the successful calibration is confirmed on the display. The
transmitter will return to the measurement mode in approximately 5 seconds.
11.2.1.6 Calibrate analog
Select the Analog Output you wish to calibrate. Each analog output can be calibrated at 4 and
20 mA.
Connect an accurate milliamp meter to the analog output terminals and then adjust the five
digit number in the display until the milliamp meter reads 4.00 mA and repeat for 20.00 mA.
As the five digit number is increased the output current increases and as the number is
decreased the output current decreases. Thus coarse changes in the output current can be
made by changing the thousands or hundreds digits and fine changes can be made by chang-
ing the tens or ones digits.
Pressing the [ENTER] key after entering both values will bring up a confirmation screen. Select-
ing No will discard the entered values, selecting Yes will make the entered values the current
ones.