Mettler Toledo M300 ISM Manual de usuario Pagina 6

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 14
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 5
17
METTLER TOLEDO US and Canada
Conductivity Standard Solutions
Provided for sensor verification and recalibration, conductivity standards are pro-
duced, analyzed and documented in the METTLER TOLEDO Thornton ISO9001
certified facility with processes similar to those used to calibrate high accuracy
Thornton conductivity sensors. They are provided with label and certificate with
lot number, certified value, expiration date, plus ASTM and NIST traceability data.
These standards are analyzed and used at equilibrium with the atmosphere.
Specifications
Standard Accuracy Shelf Life Order Number
25 µS / cm, 500 ml, HCl ± 3 % 6 months 58 078 001
100 µS / cm, 500 ml, KCl ± 1 % 12 months 58 078 002
1000 µS / cm, 500 ml, KCl ± 1 % 12 months 58 078 003
10,000 µS / cm, 500 ml,KCl ± 1 % 12 months 58 078 004
100,000 µS / cm, 500 ml, KCl ± 1 % 12 months 58 078 005
?
Did You Know
Thornton conductivity
systems are routinely used
by other instrument suppliers as the
reference to provide traceability when
calibrating their instrumentation.
Calibration of Conductivity Sensors
Thornton Auto-loop Factory Calibration System
Thornton conductivity sensors are industry standards for determining water purity
and solution concentration. Thornton ISO 9001 factory calibration and certification
are NIST and ASTM traceable using Thornton’s unique ultrapure auto-loop calibra-
tion system. Certification includes test and accuracy, plus materials as noted in
sensor specifications.
USP pharmaceutical water monitoring requirements are met with sanitary sensors
which provide accurate conductivity and temperature measurement. 316L stainless
steel Tri-Clamp mounting sensors have an electropolished finish with roughness
average (R
a
) < 0.2 µm, < 8 µin.
4-electrode sensors are ideal for monitoring high conductivity applications,
clean-in-place (CIP) solutions and deionizer regenerant concentrations.
Microelectr.
Power
Pharma Water
Wastewater
Vista de pagina 5
1 2 3 4 5 6 7 8 9 10 11 12 13 14

Comentarios a estos manuales

Sin comentarios