Specifications
Accuracy ± 10% at 25 °C, typical
Threshold settings
Restivity 50 k, 200 k, 500 k, 1 M & 2 MΩ-cm
Conductivity
20, 5, 2, 1 & 0.5μS/cm
Visual Indication
continuous green = satisfactory water quality and operation within time limit
continuous red = unacceptable water quality
flashing red = time limit (90 or 180 days) exceeded, satisfactory water quality
Power requirement
115 VAC ± 10% 50/60 Hz, 230 VAC ± 10% 50/60 Hz, or 6 VAC 50/60 Hz
Timer elapsed time is retained in memory on power loss
Installation fittings
1/2" NPT & 9/16" - 12
Operating pressure
0-150 psi (10 bar(g))
Operating temperature
+5°C to +45°C (Measurement is not temperature compensated)
Storage temperature
-
40°C to +50°C
Materials
Noryl SE 100X body, gold-plated electrodes, Viton o-ring, Lexan 121 cover
Weight
Light and transformer - 0.9 lbs (0.4 kg)
Rating
120 VAC transformer cULus listed (Canadian and US standards)
230 VAC transformer, CE compliant, European round pin plug
Ordering Information
Description Order Number
Pure Light with 120 VAC Transformer 58 001 031
Pure Light without Transformer 58 001 030
Pure Light with 230 VAC Transformer 58 001 032
Accessory
120 V Transformer 36143
Microelectr.
Power
Pharma Water
0.300
(7.6)
1.09
(27.7)
4.15
(105.3)
9/16-12 UNC
Thread
0.460 (16.7)
1/2 NPT
10 ft. (3 m)
Cable
TRANSFORMER
1.57
(40)
2.75
(70)
1.25 (32)
Wrench Flats
!
1.25
(32.74)
1.75
(44.4)
Dimensions: inches (mm), except where noted otherwise