Mettler Toledo M300 ISM Manual de usuario Pagina 12

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 14
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 11
23
METTLER TOLEDO US and Canada
Pharma Water
Ordering Information
Description Order Number
Pharma Waters UniCond
®
ISM Verifier for Thornton M800 and M300 ISM transmitters 58 082 311
Pharma Waters Verifier for analog conductivity sensors and Thornton M300 transmitter 58 082 301
Nominal Simulated Values and Resistances
Pharma Waters UniCond ISM Verifier/M800, M300ISM
Resistance/Conductivity Resistance/Temperature Nominal Simulated Value
200KΩ 0.50µS/cm
75KΩ 1.3µS/cm
47KΩ 2.1µS/cm
33KΩ 3.0µS/cm
1.10KΩ 25°C
1.33KΩ 25°C
Pharma Waters Verifiers for analog conductivity sensors/M300
Resistance/Conductivity Resistance/Temperature Nominal Simulated Value
200KΩ 0.50µS/cm
100KΩ 1.0µS/cm
75KΩ 1.3µS/cm
47KΩ 2.1µS/cm
41KΩ 2.4µS/cm
33KΩ 3.0µS/cm
1.05KΩ 13°C
1.097KΩ 25°C
1.33KΩ 85°C
?
Did You Know
The Pharma Waters Verifiers are
designed to simulate specific con-
ductivities and temperatures when used with
either an M300 transmitter or a UniCond
®
conductivity sensor and M300 ISM or M800
transmitter. The simulated conductivity and
temperature values represent values that are
in the range of conductivity limits found in
USP <645> Water Conductivity.