6
Installation examples (drawings) GPro™ 500 TDL Series
Installation examples
Dimensions of the
590 mm probe
607 (23.90”)
590 (23.23”)
100 (3.94”)
400 (15.75”)
90 (3.54”)
490 (19.29”)
3
4
2
1
½" NPT
G ¼"
G ¼"
50 (± 0,3 mm)
(19.68 [± 0.01"])
Definition of the lengths:
➊
Nominal path length, the default
length when GPro 500 is delivered.
It corresponds to the effective path
length without purging.
➋
Probe length, the physical length
of the probe.
➌
Insertion length, part of the probe
that has to protrude into the pipe
for effective purging.
➍
Effective path length, when con-
figuring the GPro 500 with the
M400, the double value of the
effective path length must be keyed
in (2⫻effective path length).