Mettler Toledo M400/2(X)H Manual de usuario

Busca en linea o descarga Manual de usuario para De medición, control y control Mettler Toledo M400/2(X)H. GPro™ 500 TDL Series In-situ sensor convenience with Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 16
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
GP
ro™
5
0
0
TDL S
eries
In-situ sensor convenience
with the power of an analyzer
T
e
c
h
n
ic
a
l d
a
ta
Overview / General operation 2
Influences on the measurement / Typical applications 3
Installation examples (drawings) 4
System overview 8
Technical specifications 9
Signal cable connections 11
Wiring diagram (ATEX and US version) 12
TDL configurator sheet 14
Ordering information 15
Short description
The GPro 500 TDL Series is designed for tough and challenging O
2
applications. The series is highly suited
to higher demanding applications where accuracy and fast response is crucial despite a varying back-
ground gas composition and a high dust load. The GPro Series can easily be calibrated without interrupting
the process,and the purging keeps the wetted parts clean to minimize maintenance.
Outstanding features are:
Interference-free in-situ measurement technology
Easy calibration without interrupting the process
12 month recommended verification interval
Different probe insertion lengths to suit applications with all types of geometry, incl. DN100 pipes
Large selection of materials for the wetted part
Easy installation with either one flange or, in short path setups, a perpendicular two flange design
No pre-alignment of flanges required etc.
Approval for hazardous areas ATEX zone 1, IECEx and FM Class 1 Div 1
Optional direct current outputs for installation without M400.
Content
GPro 500 TDL
Vista de pagina 0
1 2 3 4 5 6 ... 15 16

Indice de contenidos

Pagina 1 - GPro 500 TDL

GPro™500 TDL SeriesIn-situ sensor convenience with the power of an analyzerTechnical dataOverview / General operation 2Influences on the measurement /

Pagina 2

10Operating conditionsTemperature range process 0…+250°C (+32…+482°F)optional: 0…+600°C (+32…1112°F) with additional Thermal Barrier Ambient temperatu

Pagina 3

11Signal cable connections GPro™ 500 TDL SeriesSignal cable connectionsSignal Description Cable no. ColorPower +24V Power 24 V, 5 W 1 RedGND (Power) 2

Pagina 4

Wiring diagram (ATEX and US version) GPro™ 500 TDL Series12Active analog inputs (ATEX and US version)Wiring diagram with active analog outputs (ATEX a

Pagina 5 - (19.68 [± 0.01"])

Wiring diagram (ATEX and US version) GPro™ 500 TDL Series13Loop-powered analog inputs (ATEX and US version)Wiring diagram with loop-powered analog inp

Pagina 6

Gas analyzer GPro product keyGas analyzer GPro product key GPro™ 500 TDL Series14Gas Analyzer GPro™ 500 –YYYYYYYYYYYYYYYYYY–YY* Other configurations u

Pagina 7

Ordering information GPro™ 500 TDL Series15Ordering informationAccessories Order no.Thermal barrier 30 034 138Junction box 30 034 149Purging box for M

Pagina 8

METTLER TOLEDO Market OrganizationsSales and Service:AustraliaMettler-Toledo Ltd.220 Turner StreetPort MelbourneAUS- 3207 Melbourne/ VICPhone +61 1300

Pagina 9 - ±0.004 mA (zero scale)

2Overview/General operation GPro™ 500 TDL SeriesOverviewProcess gas analyzers are used for continuous determination of the concentration of one or mo

Pagina 10

Influences on the measurementDust loadAs long as the laser beam is able to generate a signal for the detector, the dust load of the process gases does

Pagina 11 - 2 3 6 74 5

4Installation examples (drawings) GPro™ 500 TDL SeriesInstallation examplesDimensions of the 290 mm probeDefinition of the lengths:➊Nominal path lengt

Pagina 12 - ACTIVE ANALOG INPUTS

5Installation examples (drawings) GPro™ 500 TDL SeriesInstallation examplesDimensions of the 390 mm probeDefinition of the lengths:➊Nominal path lengt

Pagina 13

6Installation examples (drawings) GPro™ 500 TDL SeriesInstallation examplesDimensions of the 590 mm probe607 (23.90”)590 (23.23”)100 (3.94”)400 (15.75

Pagina 14

7Installation examplesRequired flanges for some typical configurations➊Nominal➋Probe➌Insertion➍Effective Pipe size Number of path length length length

Pagina 15

8System overviewThe GPro Series consists of four main components:➊Sensor headThe combined transmitter and receiver unit is called the sensor head. Thi

Pagina 16

9Technical specifications GPro™ 500 TDL SeriesTechnical specificationsMeasurementEffective path length 100, 200, and 400 mm (3.94", 7.87", a